DOC

Agilent 1260 Infinity LC System Specifications

By Alvin Gordon,2014-06-29 17:37
6 views 0
Agilent 1260 Infinity LC System Specifications

    Selected System Specifications - Agilent 1260 Infinity Isocratic LC System Pump principle

    Single channel

    Flow rate range

    0.05 5 mL/min at 600 bar

    5 10 mL/min at 200 bar

    Maximum pressure

    600 bar

    Injection principle

    Flow-through design, no overfill (drawn volume equals injected volume). Autosamplers are

    available for vials only or for vials and micro-titer plates. Manual injector up to 600 bar

    available.

    Injection volumes

    0.1 100 µL, variable in 0.1 µL steps

    Column dimensions

    Length: up to 300 mm, ID: 0.05 8 mm

    Temperature range

    10 degrees below ambient to 80 ?C (or to 100 ?C with 1290 Infinity Thermostatted Column

    Compartment)

    Temperature stability

    < ? 0.15 ?C (< ? 0.05 ?C with 1290 Infinity Thermostatted Column Compartment) Detection data rate

    80 Hz for UV, 74 Hz for fluorescence, 60 Hz for evaporative light scattering detection, higher

    data rates available with 1290 Infinity detectors

    UV noise level

    As low as 0.6 uAU/cm (1260 Infinity DAD, 60 mm cell)

    Maintenance and system test

    Front access to all maintainable parts. Maintenance instructions through multimedia CD. Time for full system test (OQ/PV) < 4 hours.

    Standard GLP features

    Early Maintenance Feedback (EMF) tracks lamp burn time, usage, number of injections with

    limits and feedback messages

Extended GLP features

    (for certain systems, refer to specifications of individual modules)

    Data Recovery Card (DRC) for data security, preventing data loss in case of communication

    breakdowns.

    Radio Frequency Identification (RFID) for data traceability. RFID tags for flow cells and UV

    lamp allow unambiguous traceability of measurement conditions.

    System control

    Through local workstation software, client/server network data system and/or local handheld

    control module (1200 Infinity Series Instant Pilot). Default control: LAN / CAN.

    Agilent 1260 Infinity Quaternary LC System - Selected System Specifications

    Pump principle

    Low pressure quaternary mixing, integrated high performance degassing unit

    Flow rate range

    0.05 - 5 mL/min at 600 bar

    5 - 10 mL/min at 200 bar

    Maximum pressure

    600 bar

    Injection principle

    Flow through design, no overfill (drawn = injected volume). Vial-only autosampler and vial

    and micro-titer plate autosampler available.

    Injection volumes

    0.1 - 100 µL, variable in 0.1 µL steps, higher injection volumes optionally available

    Column dimensions

    Length: up to 300 mm, ID: 0.05 - 8 mm

    Temperature range

    10 ?C below ambient - 80 ?C

    With optional 1290 Infinity Thermostatted Column Compartment up to 100 ?C

    Temperature stability

    < ? 0.15 ?C

    With optional 1290 Infinity Thermostatted Column Compartment < ? 0.05 ?C

    Detection data rate

    80 Hz for UV, 74 Hz for fluorescence, 60 Hz for evaporative light scattering detection, higher

    data rates available with 1290 Infinity detectors

UV noise level

    As low as 0.6 uAU/cm (1260 Infinity DAD, 60 mm cell)

    Maintenance and system test

    Front access to all maintenance parts

    Maintenance instructions through multimedia CD-ROM

    Time for full system test (OQ/PV) < 4 hours

    Standard GLP features

    Early maintenance feedback-EMF

    (tracks lamp burn time, usage, number of injections with limits and feedback messages)

    Extended GLP features

    (for certain systems, refer to specifications of individual modules)

    Data Recovery Card (DRC) for data security (prevents data losses in case of communication breakdowns depends on actually chosen detector)

    Radio Frequency Identification (RFID) for data traceability. (RFID tags for flow cells and UV lamp allow unambiguous traceability of measurement conditions)

    System control

    Through local workstation software, client/server network data system and/or local handheld control module (1200 Infinity Series Instant Pilot). Default control: LAN / CAN.

Report this document

For any questions or suggestions please email
cust-service@docsford.com